Dose and image quality at CT pulmonary angiography – comparison of first and second generation dual-energy CT and 64-slice CT

posted by Ralf Bauer, M.D. | Sep 8, 2011

Objective
To compare dose and image quality of 64-slice, first and second generation dual-energy CT (DECT) for CT pulmonary angiography (CTPA).

Methods
Totally 120 patients, 30 in each group, underwent CTPA on a first generation (group 1: single-energy, 120 kV/145 mAs; group 2: DE, 140/80 kV, 70/350 mAs) or second generation dual-source DECT device (group 3: DE, 100/Sn140 kV, 120/102 mAs; group 4: DE, 80/Sn140 kV, 202/86 mAs). CTDIvol, DLP, background noise (BN), thorax diameter and attenuation in the pulmonary trunk were compared.

Results
Thorax diameter and attenuation in the pulmonary trunk did not differ significantly (p > 0.4 and >0.19) between the groups. Mean CTDIvol and DLP were significantly lower (p < 0.003) in group 4 (6.2 ± 1.6 mGy/170 ± 41 mGycm) compared to group 1 (8.5 ± 2.6 mGy/235 ± 117 mGycm), group 2 (9.2 ± 3.3 mGy/224 ± 122 mGycm) and group 3 (8.7 ± 2.8 mGy/246 ± 86 mGycm). BN was significantly lower (p < 0.0001) in group 4 (12 ± 3 HU) and group 1 (13 ± 6 HU) compared to group 3 and 2 (16 ± 6 HU and 23 ± 9).

Conclusion
The use of second generation DECT in 80/Sn140 kV configuration allows for significant dose reduction with image quality similar to 120 kV CTPA.

Authors: Bauer RW, Kramer S, Renker M, Schell B, Larson MC, Beeres M, Lehnert T, Jacobi V, Vogl TJ, Kerl JM.
Full text: Eur Radiol. 2011 May 27. [Epub ahead of print]